AcuGenix™ FastAmpli All-in-One qPCR Master Mix with UNG-V4
AcuGenix™ FastAmpli All-in-One qPCR Master Mix with UNG-V4, M2181-4, is supplied for human and animal related DNA detection.





Product overview
AcuGenix™ FastAmpli All-in-One qPCR Master Mix with UNG-V4 is a probe-based reagent specifically formulated for qualitative and quantitative Real-Time PCR applications. It incorporates a genetically engineered fast-amplification DNA polymerase capable of completing PCR amplification within 30 min.
Available versions
| Cat. no. | Product name | Formulation | Residual DNA level |
|---|---|---|---|
| M2181-4 | AcuGenix™ FastAmpli All-in-One qPCR Master Mix with UNG-V4 | Liquid (with glycerol) | Standard |
| FM2181-4 | AcuGenix™ FastAmpli Lyo-ready All-in-One qPCR Master Mix with UNG-V4 | Lyo-ready | Standard |
| M2181-4-NA | AcuGenix™ FastAmpli All-in-One qPCR Master Mix with UNG (Low DNA)-V4 | Liquid (with glycerol) | Low Residual DNA |
| FM2181-4-NA | AcuGenix™ FastAmpli Lyo-ready All-in-One qPCR Master Mix with UNG (Low DNA)-V4 | Lyo-ready | Low Residual DNA |
Features and applications
- Amplification rate is not less than 1kb/10s, which is suitable for purified DNA rapid fluorescence quantitative PCR, and the detection can be completed within 30 minutes at the fastest.
- Excellent specificity and sensitivity.
- More suitable for multiple amplification.
- Human and animal related DNA detection.
- Genomic DNA amplification and typing detection.
- Fast amplification detection.
Components
| Component | Specification/Notes |
|---|---|
| AcuGenix™ FastAmpli All-in-One qPCR Master Mix with UNG-V4 | 2× |
Performance data
1. High- and Low-Template Balance (Amplification Efficiency)
| M2181-4- Human Genomic Four-Plex Amplification | ||||
| Channel | FAM | VIC | ROX | CY5 |
|---|---|---|---|---|
| Amplification Curves | ![]() | ![]() | ![]() | ![]() |
Conclusion 1: The M2181-4 system produced consistent amplification Rn plateaus across high-, medium-, and low-concentration templates, with excellent amplification curve profiles.
2. Fast Cycling Program Test
| Standard PCR program | 50°C 2min; 95°C 5min; 45 Cycles (95°C 10s, 55°C 40s) |
| Fast PCR program | 50°C 2min; 95°C 30sec; 45 Cycles (95°C 1s, 60°C 8s) |
| M2181-4 -Standard PCR Program | ||||
| Channel | FAM | VIC | ROX | CY5 |
|---|---|---|---|---|
| Amplification Curves | ![]() | ![]() | ![]() | ![]() |
| M2181-4 - Fast PCR Program | ||||
| Channel | FAM | VIC | ROX | CY5 |
| Amplification Curves | ![]() | ![]() | ![]() | ![]() |
Ct Value Statistics
| Channel | Template Concentration (pg/μL) | Standard PCR Program | Fast PCR Program | Fast- Standard Program (ΔCt) |
|---|---|---|---|---|
| FAM | 25 | 29.45 | 29.98 | 0.53 |
| 250 | 26.26 | 26.73 | 0.47 | |
| VIC | 25 | 30.07 | 30.57 | 0.5 |
| 250 | 26.54 | 27.24 | 0.7 | |
| ROX | 25 | 31.04 | 31.42 | 0.38 |
| 250 | 28.12 | 28.44 | 0.32 | |
| CY5 | 25 | 29.17 | 29.36 | 0.19 |
| 250 | 25.82 | 25.98 | 0.16 |
Conclusion 1: M2181-4 is compatible with the fast PCR program. Across all channels, the mean amplification ΔCt for fast PCR program (25 min) versus standard PCR cycling (70 min) was less than 1.
3. Sensitivity and Precision Test
| M2181-4 - Low-Concentration ASFV Plasmid Template Amplification | ||
| Channel | FAM (Target) | VIC (Internal Control) |
|---|---|---|
| Amplification Curves | ![]() | ![]() |
| Sample | ASFV plasmid template, 3 copies/T | |||
|---|---|---|---|---|
| Target Ct Value | 37.58 | 38.51 | 37.52 | 37.53 |
| 38.64 | 37.32 | 35.71 | 36.85 | |
| 35.65 | 36.88 | 35.59 | 36.12 | |
| 38.39 | 37.53 | 36.40 | 36.95 | |
| 36.79 | 35.48 | 36.52 | 35.78 | |
| 37.86 | 36.63 | 38.33 | 36.51 | |
| Mean | 36.96 | |||
| SD | 0.95 | |||
| Ct CV% | 2.58% | |||
| Detection Rate | 100% | |||
Conclusion 1: The M2181-4 reagent achieved 100% detection of African swine fever virus (ASFV) plasmid template at the detection limit of 3 copies/T, with a Ct CV of less than 3%.
4. Freeze-Thaw Stability Test
| 0 Freeze-Thaw Cycles; 5 Freeze-Thaw Cycles; 10 Freeze-Thaw Cycles; 15 Freeze-Thaw Cycles; 20 Freeze-Thaw Cycles | ||||
| Channel | FAM | VIC | ROX | CY5 |
|---|---|---|---|---|
| Amplification Curves | ![]() | ![]() | ![]() | ![]() |
Conclusion 1: The M2181-4 premix reagent, after 5, 10, 15, and 20 freeze-thaw cycles, showed amplification performance consistent with that of the non-freeze-thawed reagent and remained stable.
5. Accelerated Stability Test
| 0 Days Accelerated; 3 Days Accelerated; 7 Days Accelerated; 14 Days Accelerated | ||||
| Channel | FAM | VIC | ROX | CY5 |
|---|---|---|---|---|
| Amplification Curves | ![]() | ![]() | ![]() | ![]() |
Conclusion 1: The M2181-4 premix reagent, after accelerated storage at 37°C for 7 and 14 days, showed amplification performance consistent with that of the non-accelerated reagent and remained stable.
Shipping and storage
| Condition | Recommendation |
|---|---|
| Shipping Conditions | -20±5°C |
| Storage Conditions | -20±5°C |
| Shelf Life | 24M |
Downloads
AcuGenix™ FastAmpli All-in-One qPCR Master Mix with UNG-V4 (M2181-4)
Instruction Manual (PDF)COA (PDF)MSDS (PDF)AcuGenix™ FastAmpli Lyo-ready All-in-One qPCR Master Mix with UNG-V4 (FM2181-4)
Instruction Manual (PDF)COA (PDF)MSDS (PDF)AcuGenix™ FastAmpli All-in-One qPCR Master Mix with UNG (Low DNA)-V4 (M2181-4-NA)
Instruction Manual (PDF)COA (PDF)MSDS (PDF)AcuGenix™ FastAmpli Lyo-ready All-in-One qPCR Master Mix with UNG (Low DNA)-V4 (FM2181-4-NA)
Instruction Manual (PDF)COA (PDF)MSDS (PDF)Need this product lyophilized, in bulk, or in a custom format?
The feasibility assessment is free of charge. Send the assay brief and target format.