Sensi Direct All-in-One qPCR Master Mix with UNG
Sensi Direct All-in-One qPCR Master Mix with UNG, MD2091, is supplied for detection of DNA viruses related to humans and animals.



Product overview
Sensi Direct All-in-One qPCR Master Mix with UNG is a specialized qPCR reagent designed for direct quantitative detection of target genes using a probe-based fluorescence detection, bypassing the nucleic acid extraction and purification steps.
Available versions
Features and applications
- Has good tolerance to inhibitors from serum, saliva, etc.
- The amplification rate is not less than 1kb/10s, suitable for rapid extraction-free fluorescence quantitative PCR, completion of detection within 30 minutes at the fastest.
- Detection of DNA viruses related to humans and animals.
- Genomic DNA amplification and typing detection.
Components
| Component | Specification/Notes |
|---|---|
| Sensi Direct All-in-One qPCR Master Mix with UNG | 2× |
Performance data
1. High/Low Template Balance
| MD2091 System - HBV Assay Amplification | ||
| Channel | FAM (Target) | VIC (Internal Control) |
|---|---|---|
| Amplification Curves | ![]() | ![]() |
| Standard Curve | ![]() | |
Conclusion 1: In the HBV assay, MD2091 produced consistent amplification-curve Rn plateaus across high-, medium-, and low-concentration templates, with an amplification efficiency of approximately 99% and an R2 coefficient of 0.998-0.999.
2. Sensitivity and Precision Test
| MD2091 - ASFV Assay - Low-Concentration ASFV Plasmid Template Amplification | |
| Channel | FAM (Target) |
|---|---|
| Amplification Curves | ![]() |
| Sample | ASFV plasmid template, 3 copies/T | |||
|---|---|---|---|---|
| Target Ct Value | 38.45 | 35.97 | 35.65 | 36.39 |
| 38.05 | 36.88 | 34.97 | 36.00 | |
| 36.32 | 36.13 | 36.12 | 36.44 | |
| 37.39 | 36.10 | 35.73 | 36.51 | |
| 36.04 | 38.12 | 34.77 | 36.06 | |
| 37.00 | 36.10 | 36.99 | 37.20 | |
| Mean | 36.47 | |||
| SD | 0.89 | |||
| Ct Value CV% | 2.44% | |||
| Detection Rate | 100% | |||
Conclusion 1: MD2091 detected the African swine fever virus (ASFV) plasmid template at the 3 copies/T detection limit with a 100% detection rate, and the Ct value CV was less than 3%.
3. Plasma and Serum Inhibitor-Tolerance Test
| MD2091 System - HBV Assay Amplification | ||
| Channel | FAM (Target) | VIC (Internal Control) |
|---|---|---|
| HBV Purified Template | ![]() | ![]() |
| HBV Plasma Template | ![]() | ![]() |
| HBV Serum Template | ![]() | ![]() |
| Template Name | Sample Concentration (IU/ml) | Equivalent Template Concentration (IU/T) | Mean Ct Value |
|---|---|---|---|
| Purified HBV Template | 104 | 235 | 28.72 |
| 103 | 23.5 | 31.90 | |
| 50 | 1.2 | 35.78 | |
| HBV Plasma Template | 1.6×104 | 100 | 30.17 |
| 1.6×103 | 10 | 33.45 | |
| 1.6×60 | 0.6 | 37.90 | |
| HBV Serum Template | 1.6×104 | 100 | 29.87 |
| 1.6×103 | 10 | 33.17 | |
| 1.6×60 | 0.6 | 38.68 |
Conclusion 1: Based on the calculated template equivalents, the amplification efficiencies of the HBV plasma/serum templates and the purified template were comparable, indicating that MD2091 had good tolerance to plasma and serum.
4. Saliva Inhibitor-Tolerance Test
| Conventional Product MD2091 | ||||
| Channel | FAM | VIC | ROX | CY5 |
|---|---|---|---|---|
| Amplification Curves | ![]() | ![]() | ![]() | ![]() |
Conclusion 1: When saliva samples were used as templates, MD2091 consistently detected the target human genomic DNA template. Compared with the conventional product, MD2091 performed better in the direct amplification of saliva samples and was highly suitable for direct amplification of DNA from saliva samples.
Shipping and storage
| Condition | Recommendation |
|---|---|
| Shipping Conditions | -20±5°C |
| Storage Conditions | -20±5°C |
| Shelf Life | 24M |