HEK293 Residual DNA Detection Kit (qPCR), BP-QN09

qPCR kit for quantitative HEK293 host-cell DNA detection.

30 fg–300 pg/µL
HEK293 calibrator range
Internal standard
Monitors extraction and amplification
UDG
Carry-over contamination control
HEK293 Residual DNA Detection Kit (qPCR)

Product overview

HEK293 Residual DNA Detection Kit (qPCR) is a 100-reaction BIORI qPCR kit for quantitative detection of HEK293 host-cell DNA in biological-product samples. It uses probe-based qPCR chemistry with an exogenous internal standard to monitor extraction and amplification. The kit includes HEK293 calibrators from 30 fg/µL to 300 pg/µL for residual DNA quantitation workflows.

Features and applications

Features
  • Targets conserved HEK293 genomic sequences using probe-based qPCR chemistry.
  • Includes an exogenous internal standard to monitor extraction and amplification.
  • Contains UDG to help prevent carryover contamination from amplification products.
  • Supplied as a 100 Reactions/kit format.
  • Includes HEK293 calibrators from 30 fg/µL to 300 pg/µL.
  • Supports quantitative residual DNA testing across biopharmaceutical process stages.
  • Designed for use with compatible real-time PCR instruments and a suitable sample nucleic acid extraction workflow.
Applications
  • Quantitative detection of HEK293 host-cell DNA.
  • Biopharmaceutical intermediate, semi-finished product, and finished product testing.
  • Host-cell DNA impurity testing during process development and quality analysis.
  • Residual DNA qPCR workflows using supported real-time PCR instruments.
  • Extraction and recovery-control workflows with a suitable nucleic acid extraction kit.

Components

Component Specification
HEK293 qPCR MIX 1.0 mL × 2 tubes
HEK293 Internal Standard 1.0 mL × 1 tube
HEK293 Negative Control 1.0 mL × 2 tubes
HEK293 Calibrator ST1 0.5 mL × 1 tube; 30 fg/µL
HEK293 Calibrator ST2 0.5 mL × 1 tube; 300 fg/µL
HEK293 Calibrator ST3 0.5 mL × 1 tube; 3 pg/µL
HEK293 Calibrator ST4 0.5 mL × 1 tube; 30 pg/µL
HEK293 Calibrator ST5 0.5 mL × 1 tube; 300 pg/µL

Performance data

Performance Data

HEK293 Residual DNA Detection Kit (qPCR) Catalog No. BP-QN09-100

A product-specific validation study evaluated analytical performance for HEK293 residual DNA. This released version presents data for linear range, limit of quantitation, and recovery.

Key Performance at a Glance

Selected performance summary for BP-QN09-100
Performance parameterSelected result
Linear range10 fg/μL to 300 pg/μL
Limit of Quantitation (LOQ)10 fg/μL
Acceptance range for recovery70%-130%

Linearity

Observed in this validation study: across the evaluated range of 10 fg/μL to 300 pg/μL, the coefficient of determination (R2) was 0.9998 and the amplification efficiency was 100.02%.

BP-QN09-100 HEK293 Residual DNA Detection Kit (qPCR): Standard curve for the HEK293 residual DNA assay across 10 fg/μL to 300 pg/μL
Standard curve for the HEK293 residual DNA assay across 10 fg/μL to 300 pg/μL

Limit of Quantitation

Observed in this validation study: at 10 fg/μL, the mean concentration was 10.87 fg/μL, with 8.36% CV and 8.73% accuracy bias across 10 replicates.

Limit-of-quantitation verification at 10 fg/μL
ReplicateConcentration valueMean concentrationCVAccuracy bias
Replicate 110.2610.878.36%8.73%
Replicate 211.43
Replicate 310.06
Replicate 49.03
Replicate 510.61
Replicate 611.28
Replicate 711.35
Replicate 810.9
Replicate 911.66
Replicate 1012.14

Recovery

HEK293 DNA at the corresponding concentrations was spiked into three actual samples. After purification using the pretreatment kit and testing with the amplification reagent, recoveries were within 70%-130%.

Sample NameMean Measured Sample Value (pg/μL)Spike Amount (pg)Mean Measured Spiked Value (pg/μL)Elution Volume (μL)Recovery
Sample 12.12E-0130.54.75E-0110086.12%
Sample 21.96E+003055.22E+00100106.94%
Sample 32.10E+0130504.93E+0110092.61%

Performance data shown here were obtained with the reference materials, extraction workflows, instruments and acceptance criteria used in BIORI’s product-specific validation study. Users should verify performance in their own sample matrix and workflow.

Shipping and storage

Condition Recommendation
Storage Store protected from light at <= -20°C. Validity period: 24 months.
Shipping Ship frozen or under validated cold-chain conditions that maintain frozen storage.
Handling Avoid repeated freeze-thaw cycles; do not exceed 10 freeze-thaw cycles.

Need this product lyophilized, in bulk, or in a custom format?

The feasibility assessment is free of charge. Send the assay brief and target format.

Talk to an expert

Not ready for a quote? Discuss feasibility, formats, host systems or validation with the specialist behind this page — no sales script.

Topic detected: your enquiry

Reply within one business day from an application scientist. For pricing, COA or samples, use Request a quote instead.

Application scientistAssay integration & formats

Tell us what you want to discuss

Four fields. We route it to the right person and reply the way you prefer.

🔒 Used only to reply to this request. No newsletters. Privacy Policy

Need pricing, a COA or samples? Request a quote →

Quick Contact

* Your data is securely protected for essential communications only. We won’t share your details or send spam emails, Privacy Policy *