CHO Residual DNA Detection Kit (qPCR), BP-QN88, quantitatively detects residual CHO host-cell DNA in bioprocess samples using probe-based qPCR with calibrators and an internal control.

CHO Residual DNA Detection Kit (qPCR), BP-QN88

USD
$0.00
BP-QN88-100

Product information published by Zhuhai Biori Biotechnology Co., Ltd.

Features
  • Probe-based qPCR assay targeting conserved CHO cell sequences.
  • Six calibrators from 3 fg/µL to 300 pg/µL support quantitative analysis.
  • Exogenous internal standard monitors extraction and amplification performance.
  • UDG enzyme helps prevent amplification-product carryover contamination.
  • 100 reactions per kit.
  • Quantification of residual CHO host-cell DNA in biopharmaceutical process samples.
  • Testing of intermediates, semi-finished products, and finished products.
  • Process development and release-support workflows using compatible real-time PCR instruments.
Component Specification
CHO qPCR MIX 1.0 mL × 2 tubes
CHO Internal Standard 1.0 mL × 1 tube
CHO Negative Control 1.0 mL × 2 tubes
CHO Calibrator ST1 0.5 mL × 1 tube; 3 fg/µL
CHO Calibrator ST2 0.5 mL × 1 tube; 30 fg/µL
CHO Calibrator ST3 0.5 mL × 1 tube; 300 fg/µL
CHO Calibrator ST4 0.5 mL × 1 tube; 3 pg/µL
CHO Calibrator ST5 0.5 mL × 1 tube; 30 pg/µL
CHO Calibrator ST6 0.5 mL × 1 tube; 300 pg/µL
Condition Recommendation
Shipping Temperature Ship frozen on dry ice or with validated frozen cold-chain packaging.
Storage Temperature ≤ -20°C, protected from light.
Shelf Life 24 months. Avoid more than 10 freeze-thaw cycles.

Performance Data

CHO Residual DNA Detection Kit (qPCR) | Catalog No. BP-QN88-100

A product-specific validation study evaluated analytical measurement performance for CHO residual DNA. This selected upload version includes only the supported linearity, quantitation, detection, and recovery results.

Key Performance at a Glance

Selected performance summary for BP-QN88-100
Performance parameter Selected result
Linear range 0.5 fg/μL to 300 pg/μL
Limit of quantitation 0.5 fg/μL
Limit of detection 0.125 fg/μL
Recovery 70%-130% acceptance range

Analytical Measurement Performance

Observed in this validation study: across the evaluated linear range of 0.5 fg/μL to 300 pg/μL, the correlation coefficient was 0.9999 and amplification efficiency was 98.69%. At 0.125 fg/μL, 30 of 30 replicates were detected.

Recovery

Recovery was evaluated at four spike levels. Observed in this validation study: recovery values in the selected table ranged from 92.69% to 112.68%.

Spike-recovery results (source Table 2)
Spike amount (fg) Replicate 1 Replicate 2 Replicate 3 Mean Recovery
3.00E+03 2.93E+01 2.49E+01 2.92E+01 2.78E+01 92.69%
3.00E+04 3.10E+02 2.90E+02 2.93E+02 2.98E+02 99.21%
3.00E+05 2.83E+03 3.13E+03 2.61E+03 2.86E+03 95.17%
3.00E+06 3.43E+04 3.64E+04 3.07E+04 3.38E+04 112.68%

Study Scope and Limitations

Results apply to the reference materials, instruments, and study conditions used in this product-specific validation study. The inconsistent source figure is not included and is not used as specificity evidence. Users should verify suitability with their own sample matrix and workflow.