Host Cell Residual DNA qPCR Kits · Cat. no. BP-QN54-100

Human Residual DNA Detection Kit (qPCR), BP-QN54

qPCR kit for quantitative human host-cell DNA detection.

30 fg–300 pg/µL
Human calibrator range
Internal standard
Monitors extraction and amplification
UDG
Carry-over contamination control
Human Residual DNA Detection Kit (qPCR)

Product overview

Human Residual DNA Detection Kit (qPCR) is a 100-reaction BIORI qPCR kit for quantitative detection of human host-cell DNA in biological-product samples. It uses probe-based qPCR chemistry with an exogenous internal standard to monitor extraction and amplification. Human calibrators from 30 fg/µL to 300 pg/µL support residual DNA quantitation in biopharmaceutical testing workflows.

Features and applications

Features
  • Targets conserved human genomic sequences using probe-based qPCR chemistry.
  • Includes an exogenous internal standard to monitor extraction and amplification.
  • Contains UDG to help prevent carryover contamination from amplification products.
  • Supplied as a 100 Reactions/kit format.
  • Includes human calibrators from 30 fg/µL to 300 pg/µL.
  • Supports quantitative host-cell DNA impurity testing across biopharmaceutical process stages.
  • Designed for compatible real-time PCR workflows with suitable sample nucleic acid extraction.
Applications
  • Quantitative detection of human host-cell DNA.
  • Biopharmaceutical intermediate, semi-finished product, and finished product testing.
  • Human residual DNA impurity testing during process development and quality analysis.
  • Residual DNA qPCR workflows using supported real-time PCR instruments.
  • Extraction and recovery-control workflows with a suitable nucleic acid extraction kit.

Components

Component Specification
Human qPCR MIX 1.0 mL × 2 tubes
Human Internal Standard 1.0 mL × 1 tube
Human Negative Control 1.0 mL × 2 tubes
Human Calibrator ST1 0.5 mL × 1 tube; 30 fg/µL
Human Calibrator ST2 0.5 mL × 1 tube; 300 fg/µL
Human Calibrator ST3 0.5 mL × 1 tube; 3 pg/µL
Human Calibrator ST4 0.5 mL × 1 tube; 30 pg/µL
Human Calibrator ST5 0.5 mL × 1 tube; 300 pg/µL

Performance data

Performance Data

Human Residual DNA Detection Kit (qPCR) Catalog No. BP-QN54-100

A product-specific validation study evaluated analytical performance for human residual DNA. This released version presents data for linear range, limit of quantitation, and recovery.

Key Performance at a Glance

Selected performance summary for BP-QN54-100
Performance parameterSelected result
Linear range10 fg/μL to 300 pg/μL
Limit of Quantitation (LOQ)10 fg/μL
Acceptance range for recovery70%-130%

Linearity

Observed in this validation study: across the evaluated range of 10 fg/μL to 300 pg/μL, the coefficient of determination (R2) was 0.9997 and the amplification efficiency was 98.467%.

BP-QN54-100 Human Residual DNA Detection Kit (qPCR): Standard curve for the human residual DNA assay across 10 fg/μL to 300 pg/μL
Standard curve for the human residual DNA assay across 10 fg/μL to 300 pg/μL

Limit of Quantitation

Observed in this validation study: at 10 fg/μL, the mean concentration was 10.84 fg/μL, with 7.92% CV and 8.47% accuracy bias across 10 replicates.

Limit-of-quantitation verification at 10 fg/μL
ReplicateConcentration valueMean concentration valueCVAccuracy bias
Replicate 110.9110.847.92%8.47%
Replicate 211.38
Replicate 310.16
Replicate 410.60
Replicate 512.40
Replicate 611.71
Replicate 710.60
Replicate 89.87
Replicate 910.23
Replicate 1010.60

Recovery

Samples were spiked with 30.5 pg, 305 pg, and 3,050 pg of human DNA (elution volume 100 μL). After purification using the pretreatment kit and testing with the amplification kit, the recovery of each simulated sample was within 70%-130%.

Sample 1Sample 2Sample 3
Replicate 12.80E-013.07E+002.73E+01
Replicate 23.38E-013.20E+002.67E+01
Replicate 33.24E-012.86E+002.69E+01
Mean (pg/μL)3.14E-013.04E+002.70E+01
Spike Amount (pg)30.53053050
Elution Volume (μL)100100100
Recovery102.95%99.67%88.52%

Performance data shown here were obtained with the reference materials, extraction workflows, instruments and acceptance criteria used in BIORI’s product-specific validation study. Users should verify performance in their own sample matrix and workflow.

Shipping and storage

Condition Recommendation
Storage Store protected from light at <= -20°C. Validity period: 24 months.
Shipping Ship frozen or under validated cold-chain conditions that maintain frozen storage.
Handling Avoid repeated freeze-thaw cycles; do not exceed 10 freeze-thaw cycles.

Need this product lyophilized, in bulk, or in a custom format?

The feasibility assessment is free of charge. Send the assay brief and target format.

Talk to an expert

Not ready for a quote? Discuss feasibility, formats, host systems or validation with the specialist behind this page — no sales script.

Topic detected: your enquiry

Reply within one business day from an application scientist. For pricing, COA or samples, use Request a quote instead.

Application scientistAssay integration & formats

Tell us what you want to discuss

Four fields. We route it to the right person and reply the way you prefer.

🔒 Used only to reply to this request. No newsletters. Privacy Policy

Need pricing, a COA or samples? Request a quote →

Quick Contact

* Your data is securely protected for essential communications only. We won’t share your details or send spam emails, Privacy Policy *